REVIEW ARTICLE
Year : 2018  |  Volume : 26  |  Issue : 3  |  Page : 123-127

Chemical items used for preparing tissue-mimicking material of wall-less flow phantom for doppler ultrasound imaging


1 Department of Medical Physics and Radiation Science, School of Physics, Univirsti Sains Malaysia, Penang, Malaysia
2 Department of Biological Sciences, School of Science, Yarmouk University, Irbid, Jordan
3 Department of Emergency, Faculty of Medicine, JUST, Irbid, Jordan

Date of Submission25-Dec-2017
Date of Acceptance20-Mar-2018
Date of Web Publication14-Sep-2018

Correspondence Address:
Prof. M Z Matjafri
Department of Medical Physics and Radiation Science, School of Physics, Univirsti Sains Malaysia, 11800 Penang
Malaysia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/JMU.JMU_13_17

Rights and Permissions
  Abstract 

The wall-less flow phantoms with recognized acoustic features (attenuation and speed of sound), interior properties, and dimensions of tissue were prepared, calibrated, and characterized of Doppler ultrasound scanning demands tissue-mimicking materials (TMMs). TMM phantoms are commercially available and ready-made for medical ultrasound applications. Furthermore, the commercial TMM phantoms are proper for ultrasound purpose or estimation of diagnostic imaging techniques according to the chemical materials used for its preparation. However, preparing a desirable TMM for wall-less flow phantom using a specific chemical material according to the specific applications is required for different flow. In this review, TMM and wall-less flow phantoms prepared using different chemical materials and methods were described. The chemical materials used in Doppler ultrasound TMM and wall-less flow phantoms fabricated over the previous decades were of high interest in this review.

Keywords: Acoustical properties, tissue-mimicking material, vessel-mimicking material, wall-less flow phantom


How to cite this article:
Oglat AA, Matjafri M Z, Suardi N, Oqlat MA, Abdelrahman MA, Oqlat AA, Farhat OF, Alkhateb BN, Abdalrheem R, Ahmad MS, Abujazar MY. Chemical items used for preparing tissue-mimicking material of wall-less flow phantom for doppler ultrasound imaging. J Med Ultrasound 2018;26:123-7

How to cite this URL:
Oglat AA, Matjafri M Z, Suardi N, Oqlat MA, Abdelrahman MA, Oqlat AA, Farhat OF, Alkhateb BN, Abdalrheem R, Ahmad MS, Abujazar MY. Chemical items used for preparing tissue-mimicking material of wall-less flow phantom for doppler ultrasound imaging. J Med Ultrasound [serial online] 2018 [cited 2023 Mar 25];26:123-7. Available from: http://www.jmuonline.org/text.asp?2018/26/3/123/231985


  Introduction Top


Since the 1960s, tissue-mimicking material (TMM) has been utilized for the preparation and characterization of ultrasound imaging. Wall-less flow phantoms are as well utilized to examine the performance of ultrasound device for practicing of sonographers. The achievement of equivalent TMM is a necessary to process for a quality monitor of Doppler ultrasound diagnostic instrument. It is essential that chemical items utilized in the TMM are prepared in a planned method to be nearly equal to the acoustical properties of real tissue with attenuation and speed of sound of 1540 ± 30 m/s, <0.5 dB/cm at MHz, respectively.[1],[2],[3]

Flow phantom is a model of TMM with a vessel-mimicking material (VMM) surrounding it during pumping of blood-mimicking fluid (BMF).[4],[5],[6],[7] The acoustical features of the different ingredients of the flow phantom correspond to the acoustical features of human blood, tissue, and vessel.[8] As required and identified by the International Electrotechnical Commission (IEC 61685 standard 1999),[10] it can be applied for a proper BMF and TMM.[10],[11],[12]

However, when the tubing materials are lacking acoustic properties, the deformation of the Doppler spectrum will lead to the refraction at the vessel wall [13],[14] and attenuation.[15],[16] Regarding acoustical and physical properties, the most proper tubing materials are known as C-flex™. The acoustic speed in the tube should be identical (the same ranges to the TMM) to prevent refraction artifacts.[17] The speed of sound in TMM is usually 1540 m/s.[17],[18],[19],[20] The refraction artifacts can be noticed when using tubes with a high velocity of sound.[21]

Several researchers [17],[19],[22],[23],[24],[25],[26],[27],[28],[29],[30] have measured and examined both the acoustic speed and attenuation of the tube (TMM) by pulse echo signal technique. Through comprehensive literature review, the studies measured and calculated the speed of sound and attenuation through the solid (TMM) samples via measuring the time of flight (ToF) or time shift, t, of the signal wave pulse from the reflection waves by applying Equations 2.1 and 2.2, respectively.[23],[31],[32]



Where SS is speed of sound of sample, l is sample thickness, depth, or distance, t is the ToF of the signal wave pulse from the reflection wave.



Where α is the attenuation coefficient of sample, −D is the difference in distance or depth of sample in mm, Ap1 is the power signal amplitude at frequency f and x, y positions of (reference signal) with no presence sample (amplitude of transmitted signal wave), Ap2 is the power signal amplitude at frequency f and x, y position through the sample (amplitude of received signal wave).

Doppler ultrasound signals were obtained in good form because the attenuation of TMM was suitable (<0.5 dB/cm/MHz) according to the IEC recommendations. Moreover, for this reason, the C-flex tubing like a VMM which has a speed of sound identical to human tissue was not used because the attenuation of C-flex is nearly ten times greater than the tissue.[9],[33] Furthermore, the attenuation of C-flex tube avoids penetrating the ultrasonic beam to the depth and get a Doppler ultrasound signal with the probe under investigation. This type of tube has an attenuation of 58 dB/cm at 8 MHz and velocity of sound of 1557 m/s.[16],[34] However, acoustic features of some tissues are shown in [Table 1].
Table 1: Acoustic features of tissues

Click here to view



  Chemicals Materials for Preparation Tissue Mimicking Material Top


Sigma silicon carbide

Sigma silicon carbide (SC) is a compound consisting of carbon and silicon. SC powder grains are joined together to form strict ceramics that are mostly used in several applications. Moreover, in this research study, SC was used as a backscatter factor to mimic real-tissue scatter properties.

Sigma aluminum oxide

Aluminum oxide (Al2O3) is a chemical construction made of both the aluminum and oxygen. It is called alkoxide, alundum, or alumina. Furthermore, Al2O3 was used in this research study for attenuation and backscatter function to mimic real-tissue properties.[16] The diameter of Al2O3 used was 3.0 and 0.3 μm.

Sigma konjac root glucomannan powder

Konjac powder consists of dietary fiber (water-soluble polysaccharide). It is a food additive utilized as a thickener and an emulsifier. Konjac powder is found in some plant sorts. However, it was used because its acoustically can be distinguished through a range of frequencies between 5 and 60 MHz and has been applied in the structure of a clinical and preclinical flow phantom.[17],[35]

Sigma carrageenan powder for gel preparation

Carrageenan powder is a gel-forming carbohydrate and is proper for emulsions and surfactants. Carrageenan powder melts at high-temperature degree >60°C. Moreover, it was used because it was of the same function of konjac powder which mentioned above (acoustically distinguished through range of frequencies between 5 and 60 MHz).[17],[35]

Sigma potassium chloride

Potassium chloride (Pc) is a mineral binary salt compound composed of potassium and chloride. It is white and odorless with crystal appearance. It is used in medicine or generally as a fertilizer for fertility the TMM.

Sigma glycerol

Glycerol is viscous fluid and mostly used in molecular biology research studies. Glycerol was used in the preparation of BMF and other viscous fluids for investigation of acoustical properties. In addition, the glycerol is an active solvent and is a popular part in pharmaceuticals. Furthermore, it is nontoxic, odorless, and colorless liquid with a sweet taste. The molecular weight of glycerol is 92.1 g/mol and molecular formula is C3H8O3. Furthermore, the glycerol density is 1.25 g/mL. Glycerol is used as a viscous fluids for the investigation of acoustical properties in mixture fluid of BMF used in this research.

Sigma gelatin

Gelatin is a diverse (heterogeneous) mixture of a large molecular mass of the water-soluble proteins. It exists in collagen and has a density of about 1.3 g/ml. It is used to improve attachment and strength of many items. Gelatin utilized as a thickener and stabilizer in industrial and engineering applications. It was used in this project as a material to increase the strength of TMM at high flow and to keep the acoustical properties suitable.

Among the earliest TMMs prepared for ultrasound imaging were gelatin-based materials. Gelatin, a homogeneous colloid gel, is primarily derived from collagen in animal tissues. The Madsen group mixed gelatin with varying concentrations of alcohol and uniformly distributed graphite powder, with p-methyl and p-propyl benzoic acid used as preservatives against bacterial invasion.[36],[37] Depending on the concentration of n-propanol in water, a speed of sound between 1520 and 1650 m/s at room temperature could be achieved. By varying the concentration of graphite powder in the gelatin-based compound, therefore, adjusting the scattering coefficient, the attenuation coefficient was varied between 0.2 and 1.5 dB/cm at 1 MHz. These materials were reported to have high stability near-room temperature over a period of 4 months, provided that the samples were stored in a closed container below a layer of distilled water. Reported disadvantages to this technique were instability with temperature variations, susceptibility to microbial invasion, and difficulty in achieving uniform distribution of graphite scatters as the particles settled during cooling.[38],[39],[40],[41],[42]

A summary for each chemical with their purpose for preparing the TMM in this review was shown in [Table 2] and [Figure 1].
Table 2: A brief summary for each chemical with their purpose for preparing the tissue-mimicking materials

Click here to view
Figure 1: Chemicals items used for tissue-mimicking material flow phantom fabrication: (a) silicon carbide, (b) aluminum oxide, (c) K, (d) C, (e) potassium chloride, (f) glycerol, (g) gelatin

Click here to view



  Tissue-Mimicking Materials for Wall-Less Flow Phantoms Top


There are several factors considered with fabrication: a wall-less flow phantom should be suitable with various applications, such as depth, flow waveform (straight, tortuous, oblique), flow rate, and sizes (diameters) of the vessel. For clinical application, the ideal vessel diameters between 3 and 10 mm are recommended and should mimic human vessels. For preclinical uses, the vessel diameters are ideally between 0.5 and 2 mm and should mimic rat and mouse vessels. For clinical application, the highest phantom dimension is ideally between 150 and 300 mm; however, for preclinical application, the range is between 50 and 100 mm. The dimensions of preclinical flow phantom are usually less than that of clinical flow phantom.[17],[43]

VMM is composed of both silicone (polydimethylsiloxane) and elastomer (Sylgard 184). The setup is by integrating a base and a treating factor permitting the mixture to be treated for 7 days at specific temperatures ranging between 25°C and 35°C. Four mixtures of cellulose particles with 50 μm as a scatters with different concentrations by weight (0%, 1%, 3%, and 5%) were added and examined for relevant properties, including the speed of sound, visual appearance, and attenuation with the B-mode ultrasound image. Furthermore, the TMM used a different material mixed with the specific amount by weight such as (SC, Al2O3 powder, formaldehyde, high-gel strength agar, glycerol, and distilled water) to obtain a proper backscatter and attenuation properties. The pulse-transmission mechanism used to measure the TMM attenuation coefficients four Sylgard 184 mixtures. All measurement were conducted via single-element unfocussed probe and pulse extend from 2 to 7.5 MHz at room temperature.[44],[45]

There is no tubing material applied to fabricate the wall-less flow phantoms, and the BMF is an immediate link with the TMM. Fabrication of wall-less flow phantoms is necessary to prevent problems that resulted by US deformation through the tube wall.[15],[46] TMM was created by utilizing two hydrogel materials, namely konjac and carrageenan (KC). The KC is composed of two organic materials and was added with water to produce a flexible and strong material. TMM is prepared by mixing SC, Al2O3, glycerol, and KC materials. The KC-based TMM had a velocity of sound of 1548 ± 3 m/s and an attenuation (dB/cm) behavior of 0.01024f2 + 0.3639f, where f is transferred frequency (MHz).[35],[47]

The wall-less flow phantoms made of metal cores containers with dimensions of 10 cm × 10 cm × 23 cm were mounted in melting temperature of 69°C. Before pouring, the phantom was fixed by reticulated foam and put around the connectors. TMM was composed of agar, then poured into the container, and let it set. However, used of agar-based TMM failure was due to leakage the BMF with high flow rate. The problem was solved via the expansion of a new powerful TMM relied on utilization of two types of gels; KC was used instead of agar. The ingredients by (%weight) were konjac powder, carrageenan powder, Al2O3 powder 3 μm, Al2O3 powder 0.3 μm, H2O, 400 grains SC powder, glycerol, and Pc, which used to adjusting the temperature of the gel with %weight 1.5%, 1.5%, 0.96%, 0.89%, 84%, 0.54%, 10%, and 0.7%, respectively. Several items were estimated as a portion of the European Commission project which made the TMM.[22],[48] The acoustical properties of TMM were perfect, and it is meeting the requirements of IEC 1685 draft report: At room temperature, it had a velocity of sound of 1550 ± 6 m/s and an attenuation (dB/cm) behavior of 2.81 ± 0.08 with 5 MHz as a frequency. The brand of ultrasound scanner equipment used for examining the flow phantom was named HDI 5000 Philips Medical ultrasound system. The system is provided with linear array transducer L12-5 to scan the flow channels and collect longitudinal imaging of all areas and measurement of flow velocity.[35],[49]

Another wall-less flow phantom was made of 9.52 mm as a vessel diameter. The TMM made of a mixture of 8% glycerol, 3% of very strong agar gel, and 89% of distilled water has the acoustic speed identical to that of soft tissue (ST). The metal rod worked like the mold for the vessel. Vessel lumen preserved in its original form throughout time and the wall less flow phantom was not absorbing water. To prevent drying of the agar gel and potential deformation, the phantom was maintained in a water bath. Experiment research was carried out at room temperature in a horizontal constant flow loop sample. Linear array transducer with 4-MHz (L7-4) was utilized to generate Doppler scan images.[50]

The great-strength Struers agar is utilized because it produces TMM with more firm and robust, and thus, the likelihood of breakage or tear will decrease when TMM is exposed to the strong high pressure of high flow rates. The primary function of Al2O3 powder is to influence the attenuation with a linear proportional of frequency. Furthermore, Al2O3 and SC powder adjust the backscatter to mimic tissue. Benzalkonium chloride is used to avoid the development of microorganisms, and glycerol is utilized to supply the required speed of sound. TMM prepared by mixing SC, Al2O3, benzalkonium chloride, glycerol, and KC materials continuously with a magnetic stirring rod. It was then cool to 42°C with keep of stirring and then poured the mixing materials into the test container. The acoustical properties of TMM were perfect, and it is meeting the requirements of IEC 1685 draft report: At room temperature, it had a velocity of sound of 1541 ± 3 m/s and an attenuation (dB/cm) behavior of 0.5 ± 0.03 with nearly 3–10 MHz as a frequency. The VMM made by casting the metallic rod inside the wall-less TMM channel with about 8-mm diameter. The brand of ultrasound scanner equipment which used for examining this flow phantom was named HDI 5000 ATL ultrasound system. This system provided a linear array transducer to scan the flow channels and doing a measurement of flow velocity.[16]


  Conclusion Top


At present, even though the TMMs are abundant commercially, customized TMM keeps. Many soft TMMs have been described that have compressional acoustical and physical features within the measured range of STs. The backscattering coefficient, nonlinearity parameter, and shear wave speed of sound (in the case of hard tissue substitutes) have rarely been reported.

Several soft TMMs that have different acoustical and physical properties (density, attenuation, and speed of sound) have been explained. The inclusive information steps for the preparation of TMM and wall-less flow phantom items in this review article can be useful to medical and engineering researchers, particularly, in the knowledge to make TMM and wall-less flow phantom proper fabricated for Doppler ultrasound applications. At present, even the TMM are abundant commercially. However, customized TMM keeps having a function, first in the medical Doppler ultrasound research group. The main advantage of TMM is the low-cost and proper elasticity of chemical items.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Kondo T, Fujimoto H. Ultrasound tissue-mimicking materials using oil gel and measurement of their characteristics. Jpn J Appl Physics 2002;41:3598.  Back to cited text no. 1
    
2.
Oglat AA, Matjafri MZ, Suardi N, Abdelrahman MA, Oqlat MA, Oqlat AA. Anew scatter particle and mixture fluid for preparing blood mimicking fluid for wall-less flow phantom. J Med Ultrasound 2018; [In press].  Back to cited text no. 2
    
3.
Oglat AA, Suardi N, Matjafri Mz, Oqlat MA, Abdelrahman MA, Oqlat AA. A review of suspension-scattered particles used in blood-mimicking fluid for Doppler ultrasound imaging. J Med Ultrasound 2018; [In press].  Back to cited text no. 3
    
4.
Hoskins P, Anderson T, McDicken W. A computer controlled flow phantom for generation of physiological Doppler waveforms. Physics Med Biol 1989;34:1709.  Back to cited text no. 4
    
5.
Law Y, Johnston K, Bascom P. Computer-controlled pulsatile pump system for physiological flow simulation. Med Biol Eng Comput 1987;25:590-5.  Back to cited text no. 5
    
6.
Browne JE. A review of doppler ultrasound quality assurance protocols and test devices. Phys Med 2014;30:742-51.  Back to cited text no. 6
    
7.
Lai SS, Yiu BY, Poon AK, Yu AC. Design of anthropomorphic flow phantoms based on rapid prototyping of compliant vessel geometries. Ultrasound Med Biol 2013;39:1654-64.  Back to cited text no. 7
    
8.
Oglat AA, Matjafri M, Suardi N, Oqlat MA, Oqlat AA, Abdelrahman MA. A new blood mimicking fluid using propylene glycol and their properties for a flow phantom test of medical doppler ultrasound. International Journal of Chemistry, Pharmacy & Technology 2017;2:220-31.  Back to cited text no. 8
    
9.
Baghdar HN, Nematy M, Kooshyar MM, Taghipour A, Tabassi SA, Shokri S, et al. Effect of a functional food (vegetable soup) on blood rheology in patients with polycythemia. Avicenna Journal of Phytomedicine 2018:1-11.  Back to cited text no. 9
    
10.
Wells Jr RE, Merrill EW. Influence of flow properties of blood upon viscosity-hematocrit relationships. J Clin Invest 1962;41:1591.  Back to cited text no. 10
    
11.
Baskurt OK, Meiselman HJ. Blood rheology and hemodynamics. in Seminars in thrombosis and hemostasis. New York, NY 10001, USA: Thieme Medical Publishers, Inc.; 2003.  Back to cited text no. 11
    
12.
Baskurt OK, Meiselman HJ. Blood rheology and hemodynamics in Seminars in thrombosis and hemostasis. New York, NY, USA: Thieme Medical Publishers, Inc.; 2003.  Back to cited text no. 12
    
13.
Thompson R, Aldis G, Linnett I. Doppler ultrasound spectral power density distribution: Measurement artefacts in steady flow. Med Biol Eng Comput 1990;28:60-6.  Back to cited text no. 13
    
14.
Oglat, A.A., M. Matjafri, N. Suardi, M.A. Oqlat, M.A. Abdelrahman, and A.A. Oqlat, A review of medical doppler ultrasonography of blood flow in general and especially in common carotid artery. Journal of Medical Ultrasound, 2018;26: p. 3.  Back to cited text no. 14
    
15.
Rickey DW, Picot PA, Christopher DA, Fenster A. A wall-less vessel phantom for doppler ultrasound studies. Ultrasound Med Biol 1995;21:1163-76.  Back to cited text no. 15
    
16.
Ramnarine KV, Anderson T, Hoskins PR. Construction and geometric stability of physiological flow rate wall-less stenosis phantoms. Ultrasound Med Biol 2001;27:245-50.  Back to cited text no. 16
    
17.
Zhou X, Kenwright DA, Wang S, Hossack JA, Hoskins PR. Fabrication of two flow phantoms for doppler ultrasound imaging. IEEE Trans Ultrason Ferroelectr Freq Control 2017;64:53-65.  Back to cited text no. 17
    
18.
Browne JE, Watson AJ, Hoskins PR, Elliott AT. Validation of a sensitivity performance index test protocol and evaluation of colour doppler sensitivity for a range of ultrasound scanners. Ultrasound Med Biol 2004;30:1475-83.  Back to cited text no. 18
    
19.
Yoshida T, Sato K, Kondo T. Blood-mimicking fluid using glycols aqueous solution and their physical properties. Jpn J Appl Physics 2014;53:7KF01.  Back to cited text no. 19
    
20.
Kenwright DA, Laverick N, Anderson T, Moran CM, Hoskins PR. Wall-less flow phantom for high-frequency ultrasound applications. Ultrasound Med Biol 2015;41:890-7.  Back to cited text no. 20
    
21.
Sato M, Ishida H, Konno K, Komatsuda T, Furukawa K, Yamada M, et al. Analysis of refractive artifacts by reconstructed three-dimensional ultrasound imaging. J Med Ultrason 2006;33:11-6.  Back to cited text no. 21
    
22.
Ramnarine KV, Nassiri DK, Hoskins PR, Lubbers J. Validation of a new blood-mimicking fluid for use in doppler flow test objects. Ultrasound Med Biol 1998;24:451-9.  Back to cited text no. 22
    
23.
Samavat H, Evans J. An ideal blood mimicking fluid for doppler ultrasound phantoms. J Med Physics 2006;31:275.  Back to cited text no. 23
    
24.
Thorne ML, Poepping TL, Rankin RN, Steinman DA, Holdsworth DW. Use of an ultrasound blood-mimicking fluid for doppler investigations of turbulence in vitro. Ultrasound Med Biol 2008;34:1163-73.  Back to cited text no. 24
    
25.
Oates CP. Towards an ideal blood analogue for doppler ultrasound phantoms. Phys Med Biol 1991;36:1433-42.  Back to cited text no. 25
    
26.
Yang P, Zhu H. Influence of Transducer Focus Position and Signal Length in Backscatter Coefficient Measurement for Blood Mimicking Fluid. In Biomedical Engineering and Computer Science (ICBECS), 2010 International Conference on. 2010. IEEE.  Back to cited text no. 26
    
27.
Kimme-Smith C, Hussain R, Duerinckx A, Tessler F, Grant E. Assurance of consistent peak-velocity measurements with a variety of duplex Doppler instruments. Radiology 1990;177:265-72.  Back to cited text no. 27
    
28.
Boote EJ, Zagzebski JA. Performance tests of doppler ultrasound equipment with a tissue and blood-mimicking phantom. J Ultrasound Med 1988;7:137-47.  Back to cited text no. 28
    
29.
Tanaka K, Yoshida T, Sato K, Kondo T, Yasukawa K, Miyamoto N, et al. Blood-Mimicking fluid for testing ultrasonic diagnostic instrument. Jpn J Appl Physics 2012;51(7S):07GF18.  Back to cited text no. 29
    
30.
Hoskins P, Loupas T, McDicken W. A comparison of the Doppler spectra from human blood and artificial blood used in a flow phantom. Ultrasound Med Biol 1990;16:141-7.  Back to cited text no. 30
    
31.
Halim MA, Buniyamin N, Mohamad Z. Improving intramuscular fat measurement by considering the thickness of protective layer in ultrasonic transducer. In Electrical, Electronics and System Engineering (ICEESE), 2014 International Conference on. 2014. IEEE.  Back to cited text no. 31
    
32.
Halim MA, Buniyamin N, Shari MM, Kassim RA. Ultrasound velocity to measure beef, chicken and fish fillet fat content. Research and Development (SCOReD), IEEE Student Conference on 2012. IEEE; 2012.  Back to cited text no. 32
    
33.
Hoskins PR. Simulation and validation of arterial ultrasound imaging and blood flow. Ultrasound Med Biol 2008;34:693-717.  Back to cited text no. 33
    
34.
Hoskins P, Ramnarine K. Doppler test devices. Doppler Ultrasound: Physics, Instrumentation and Signal Processing. 2nd ed., Vol. 4. Chichester, UK: Wiley; 2000. p. 2010.  Back to cited text no. 34
    
35.
Meagher S, Poepping T, Ramnarine K, Black R, Hoskins P. Anatomical flow phantoms of the nonplanar carotid bifurcation, part II: experimental validation with Doppler ultrasound. Ultrasound Med Biol 2007;33:303-10.  Back to cited text no. 35
    
36.
Madsen EL, Berg WA, Mendelson EB, Frank GR. Anthropomorphic breast phantoms for qualification of investigators for ACRIN Protocol 6666. Radiology 2006;239:869-74.  Back to cited text no. 36
    
37.
Madsen EL, Frank GR, Dong F. Liquid or solid ultrasonically tissue-mimicking materials with very low scatter. Ultrasound Med Biol 1998;24:535-42.  Back to cited text no. 37
    
38.
Madsen EL, Frank GR, Krouskop TA, Varghese T, Kallel F, Ophir J, et al. Tissue-mimicking oil-in-gelatin dispersions for use in heterogeneous elastography phantoms. Ultrason Imaging 2003;25:17-38.  Back to cited text no. 38
    
39.
Madsen EL, Hobson MA, Frank GR, Shi H, Jiang J, Hall TJ, et al. Anthropomorphic breast phantoms for testing elastography systems. Ultrasound Med Biol 2006;32:857-74.  Back to cited text no. 39
    
40.
Madsen EL, Zagzebski JA, Banjavie RA, Jutila RE. Tissue mimicking materials for ultrasound phantoms. Med Phys 1978;5:391-4.  Back to cited text no. 40
    
41.
Madsen EL, Zagzebski JA, Frank GR. An anthropomorphic ultrasound breast phantom containing intermediate-sized scatterers. Ultrasound Med Biol 1982;8:381-92.  Back to cited text no. 41
    
42.
Madsen EL, Zagzebski JA, Ghilardi-Netto T. An anthropomorphic torso section phantom for ultrasonic imaging. Med Physics 1980;7:43-50.  Back to cited text no. 42
    
43.
Kenwright DA, Laverick N, Anderson T, Moran CM, Hoskins PR. Wall-less flow phantom for high-frequency ultrasound applications. Ultrasound Med Biol 2015;41:890-7.  Back to cited text no. 43
    
44.
Poepping TL, Nikolov HN, Thorne ML, Holdsworth DW. A thin-walled carotid vessel phantom for Doppler ultrasound flow studies. Ultrasound Med Biol 2004;30:1067-78.  Back to cited text no. 44
    
45.
Culjat MO, Goldenberg D, Tewari P, Singh RS. A review of tissue substitutes for ultrasound imaging. Ultrasound Med Biol 2010;36:861-73.  Back to cited text no. 45
    
46.
Patterson M, Foster F. The improvement and quantitative assessment of B-mode images produced by an annular array/cone hybrid. Ultrasonic Imaging 1983;5:195-213.  Back to cited text no. 46
    
47.
Zhou X, Kenwright DA, Wang S, Hossack JA, Hoskins PR. Fabrication of two flow phantoms for Doppler ultrasound imaging. Vol. 64. IEEE Transactions On Ultrasonics, Ferroelectrics, and Frequency Control; 2017. p. 53-65.  Back to cited text no. 47
    
48.
Teirlinck CJ, Bezemer RA, Kollmann C, Lubbers J, Hoskins PR, Ramnarine KV, et al. Development of an example flow test object and comparison of five of these test objects, constructed in various laboratories. Ultrasonics 1998;36:653-60.  Back to cited text no. 48
    
49.
Watts DM, Sutcliffe C, Morgan R, Meagher S, Wardlaw J, Connell M, et al. Anatomical flow phantoms of the nonplanar carotid bifurcation, part I: computer-aided design and fabrication. Ultrasound Med Biol 2007;33:296-302.  Back to cited text no. 49
    
50.
Allard L, Cloutier G. Power Doppler ultrasound scan imaging of the level of red blood cell aggregation: An in vitro study. J Vasc Surg 1999;30:157-68.  Back to cited text no. 50
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2]


This article has been cited by
1 Characterization of a fat tissue mimicking material for high intensity focused ultrasound applications
Antria Filippou, Irene Louca, Christakis Damianou
Journal of Ultrasound. 2022;
[Pubmed] | [DOI]
2 Flexible lead-free piezoelectric arrays for high-efficiency wireless ultrasonic energy transfer and communication
Laiming Jiang, Bo Wu, Xiaowei Wei, Xiang Lv, Haoyue Xue, Gengxi Lu, Yushun Zeng, Jie Xing, Wenjuan Wu, Jiagang Wu
Materials Horizons. 2022;
[Pubmed] | [DOI]
3 A KNN composite-based piezoelectric helix for ultrasonic transcutaneous energy harvesting
Laiming Jiang, Haoyue Xue, Ruichen Li, Jiagang Wu
Applied Physics Letters. 2022; 120(23): 233504
[Pubmed] | [DOI]
4 Characterizing Viscoelastic Polyvinyl Alcohol Phantoms for Ultrasound Elastography
Ananya Sharma, Sai Geetha Marapureddy, Abhijit Paul, Sapna R. Bisht, Manik Kakkar, Prachi Thareja, Karla P. Mercado-Shekhar
Ultrasound in Medicine & Biology. 2022;
[Pubmed] | [DOI]
5 Transcranial Doppler Ultrasonography as a Diagnostic Tool for Cerebrovascular Disorders
Yuanmei Pan, Wenbin Wan, Minjie Xiang, Yangtai Guan
Frontiers in Human Neuroscience. 2022; 16
[Pubmed] | [DOI]
6 Preparation and Evaluation of Polymer-Based Ultrasound Gel and Its Application in Ultrasonography
Sadia Afzal, Muhammad Zahid, Zulfiqar Ahmad Rehan, H. M. Fayzan Shakir, Hamza Javed, Meshari M. H. Aljohani, Syed Khalid Mustafa, Maqsood Ahmad, Montaser M. Hassan
Gels. 2022; 8(1): 42
[Pubmed] | [DOI]
7 Multichannel Piezo-Ultrasound Implant with Hybrid Waterborne Acoustic Metastructure for Selective Wireless Energy Transfer at Megahertz Frequencies
Laiming Jiang,Gengxi Lu,Yang Yang,Yang Xu,Fangjie Qi,Jiapu Li,Benpeng Zhu,Yong Chen
Advanced Materials. 2021; : 2104251
[Pubmed] | [DOI]
8 A 10-Year Retrospective Review of Prenatal Applications, Current Challenges and Future Prospects of Three-Dimensional Sonoangiography
Tuangsit Wataganara,Thanapa Rekhawasin,Nalat Sompagdee,Sommai Viboonchart,Nisarat Phithakwatchara,Katika Nawapun
Diagnostics. 2021; 11(8): 1511
[Pubmed] | [DOI]
9 Hepatocellular Carcinoma Liver Dynamic Phantom For Mri
Muntaser S. Ahmad,Osama Makhamrah,Nursakinah Suardi,Ahmad Shukri,Nik Noor Ashikin Nik Ab Razak,Ammar A. Oglat,Hjouj Mohammad
Radiation Physics and Chemistry. 2021; : 109632
[Pubmed] | [DOI]
10 Dynamic Hepatocellular Carcinoma Model Within a Liver Phantom for Multimodality Imaging
Muntaser S. Ahmad,Nursakinah Suardi,Ahmad Shukri,Nik Noor Ashikin Nik Ab Razak,Ammar A. Oglat,Osama Makhamrah,Hjouj Mohammad
European Journal of Radiology Open. 2020; 7: 100257
[Pubmed] | [DOI]
11 Artifacts in Diagnostic Ultrasonography
Ammar A. Oglat,Marwan Alshipli,Mohannad Adel Sayah,Muntaser S. Ahmad
Journal for Vascular Ultrasound. 2020; 44(4): 212
[Pubmed] | [DOI]
12 A realistic flow phantom model of the carotid artery in preterm infants for training and research
Sujith Pereira,Jonathan Reeves,Malcolm Birch,Sakthi Finton-James,Komal Verma,Robert Krug,Ajay Sinha,Stephen Kempley
Ultrasound. 2020; 28(3): 145
[Pubmed] | [DOI]
13 Bimodal microwave and ultrasound phantoms for non-invasive clinical imaging
Enrique Villa,Natalia Arteaga-Marrero,Javier González-Fernández,Juan Ruiz-Alzola
Scientific Reports. 2020; 10(1)
[Pubmed] | [DOI]
14 Development and Characterization of Medical Phantoms for Ultrasound Imaging Based on Customizable and Mouldable Polyvinyl Alcohol Cryogel–Based Materials and 3-D Printing: Application to High-Frequency Cranial Ultrasonography in Infants
Luis Elvira,Carmen Durán,Ricardo T. Higuti,Marcelo M. Tiago,Alberto Ibáñez,Montserrat Parrilla,Eva Valverde,Javier Jiménez,Quique Bassat
Ultrasound in Medicine & Biology. 2019;
[Pubmed] | [DOI]
15 A review of suspension-Scattered particles used in blood-Mimicking fluid for doppler ultrasound imaging
AmmarA Oglat,Nursakinah Suardi,MZ Matjafri,MohammadA Oqlat,MostafaA Abdelrahman,AhmadA Oqlat
Journal of Medical Ultrasound. 2018; 26(2): 68
[Pubmed] | [DOI]
16 A new scatter particle and mixture fluid for preparing blood mimicking fluid for wall-less flow phantom
AmmarA Oglat,MZ Matjafri,Nursakinah Suardi,MostafaA Abdelrahman,MohammadA Oqlat,AhmadA Oqlat
Journal of Medical Ultrasound. 2018; 26(3): 134
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
Abstract
Introduction
Chemicals Materi...
Tissue-Mimicking...
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed6425    
    Printed325    
    Emailed0    
    PDF Downloaded680    
    Comments [Add]    
    Cited by others 16    

Recommend this journal